BENEFUSION SP5 SYRINGE PUMP
S-A-S16-SP5
BeneFusion series pump to the high-end infusion pump market, the small, lightweight and portable pumps are ideal infusion system.
The smart BeneFusion SP5 pump helps to administered anesthesia drugs, so users can concentrate on improving patient care during surgeries, it is a great partner for Anesthesia infusion!
Technical Data
Physical specification
- WxHxD: 295x87x174(mm)
- Net weight: <2.5kg, Stuckable
- Classification: Type CF, Defibrillation-proof, class 1
Display
- Screen: 3.5" TFT color LCD, 16:9 format
- Brightness: 1-8 levels adjustable, default is level 4
- Information: Delivery rate, current infusion, VTBI, total volume, syringe size, syringe brand, pressure limit, battery capacity, drugs, remaining time, alarms
Battery : 10 hours operating Time
CPU type
Syringe selection
- 5,10,20,30,50/60ml
- Pre-configured 3 syringe brands, including BD, B.Braun, Terumo
- User-defined syringe brands and quantity possible by PC tool, Brand 1 and Brand 2 as default.
- Automatic recognition of syringe size and fixation
Basic parameters
- Mode: 8 modes: Rate Mode, Time mode, body weight mode, Ramp up/down mode, Sequential mode, Loading dose mode, Micro-infusion mode, TIVA mode
- Flow rate: 5ml:0.1-100ml/h,10m:0.1-200ml/h, 20ml:0.1-400ml/h, 30ml:0.1-600ml/h,50/60ml: 0.1-1500ml/h
- Increment:0.01ml(0.1-99.99ml/h), 0.1ml(100-999.99-ml/h),1ml(1000-1500ml/h)
- Present volume(VTBI): 0.10-9999.99ml(increment 0.01ml)
- Present time: 00:00:01-99:59:59 (h:m:s)
- Accumulated volume: 0.01-9999.99ml(increment 0.01ml)
- KVO: 0.1-5.0nl/h adjustable, step 0.1ml/h, default is 1.0 ml/h
- Purge: 0.10-1500ml/h (depending on syringe size)
Bolus
- Bolus rate: Automatic/manual bolus: 0.10-1500ml/h(depending on syringe size)
- Present bolus volume: Present automatic Bolus Volume: 0.10-9999.99ml
Function spectrum
- Automatic calculation of the delivery rate based on different modes
- Infusion pause/standby: present parameters remain stored
- Drug library up to 2000 drugs,"ON" amd "OFF" selectable; use-defined drug list available
- History rcodsup to 2000, data could be transmitted to PC and saved as EXCEL format to print out.
- Titration function: change flow rate during infusion with increment 0.01ml/h
Safety system
- Occlusion pressure: 4 pressure unit: mmHg/kPa/bar/psi selectable; auto convertion in different units, default is mmHg, 11 levels occlusion pressure, 150-975mmHg selectable, default is 450mmHg
- Auto data lock: "ON" and "OFF", "ON": 1-5min.selectable, step 1 min, default is "OFF"
- Anti-bolus: Automatic reduction of the bolus volume following pressure alarm
Alarm
- Alarm type: Visual and acoustic alarms, High: occlusion, battey empty, VTBI done, syringe empty, syringe disengaed, KVO finish, system error, Middle: Standby time expired system abnormal, Low: syringe near empty, reminder, no battery, battery low, VTBI near done, no AC
- Alarm volume: 1-8 levels selectable, default is level 4
- Pre-alarms: 1-30min. before syringe is empty, 3 min. before battey is completed, 30min. as low battery
- Reminder: "ON" and "OFF", "ON": 1-5min. selectable, step 1 min.
Interface
- Rs232
- Nurse call connector
- DC adapter connector
Power supply
- 100-240V~, 50/60Hz, 0.53-0.28A; DC Voltage: 10V-16V, 2.25A-1.5A
Environmental requiremet
- Temperature: 5~40oC for operating, 20~60oC for storage
- Humidity:15~95% for operaing, 10~95% for storage
- Air pressure: 57~106kPa for operating, 50~106kPa for storage
- Water-proof: IP23
Lifetime
- >5 years under continuous duty conditions
Modes specification
- Flow rate mode: VTBI: 0.10-9999.99ml; Rate:5ml: 0.1-100ml/h; 10ml: 0.1-200ml/h; 20ml: 0.1-400ml/h; 30ml:0.1-600ml/h;50/60ml: 0.1-1500ml/h
- Time mode: VTBI: 0.10-9999.99ml; Time: 00:00:01-99:59:59 (h:m:s)
- Body weight mode: Weight: 0.1-300.0kg or 0.1-660.8lb, kg/lb selectable, Drug-Amount: 0.1-999.9,g/mg/ug,ng,lU selectable, default is mg; volume: 0.10-9999.99ml; Dose: 0.01-999.99; Dose unit: mg,ug,ng,lU/kg/h;mg,ug,ng,lU/kg/min selectable, default is mg/kg/h; VTBI: 0.10-9999.99ml; Time: 00:00:01-99:59:59(h:m:s) (Auto. Calculated)
- Ramp up/down mode: VTBI: 0.10-9999.99ml; Time range: 00:00:01-99:59:59(h:m:s)
- Sequential mode: VTBI:0.10-9999.99ml; Rate:0.10-1500ml/h; Time: 00:00:01-99:59:59(h:m:s)
- Loading dose mode: VTBI: 0.10-9999.99ml; Rate: 0.10-1500ml/h; Time: 00:00:01-99:59:59(h:m:s)
- Micro-infusion mode: VTBI: 0.10-1000.00ml;Rate:0.1-100ml/h; Time: 00:00:01-99:59:59; Bolus rate: 0.1-100ml/h
- TIVA mode: Weight: 0.1-300.0kg or 0.1-660.8lb, kg/lb selectable; Drug-Amount: 0.1-999.9, g/mg/ug/ng selectable; Volume: 0.10-9999.99ml; Dose: 0.01-999.99, Induction dose: mg/kg, ug/kg, ng/kg selectable; Maintenace rate: mg.ug.ng/kg/h; mg/ug/ng/kg/min; Induction Time: 00:00:01-99:59:59(h:m:s)
Brand: MINDRAY
Packing: 1 UNIT