Specification:
MODEL | HK-100II |
Infusion Mode | Rate Mode, Drip Mode, Time Mode, Weight Mode, Dose Mode, TPN Mode, Program Mode, Intermittent Mode, Micro Mode, Sequential Mode, Drug Library |
Applicable infusion set | Applicable infusion set |
Applicable infusion set | ±5% |
Infusion rate | (0.10-1200)ml/h,increment at 0.01ml/h |
VTBI | 0-9999.99)ml |
Volume infused | 0-36000 ml |
KVO rate | (0,0.10~10.0)ml/h |
BOLUS rate | (0.10-1200)ml/h |
Purge rate | (0.10~1200)ml/h,increment at 0.01ml/h |
Bubble detection | Ultrasound sensor detection with 4 levels :50ul-300ul |
Occlusion pressure | 13 Level |
History record | 1300 |
Alarm | Use battery, AC Fail, Door open, Occlusion, Almost done, Near empty, Drip Abnormal, No operation, Infusion completion, Air Bubble, Low Battery, Battery exhausted, malfunction etc |
Features | Anti-Bolus,Anti-reverse detection,Double CPU,Key lock,Auto shutdown |
Optional Functions | Nurse call,Wifi,Drop sensor |
Waterproof | IP24 |
Power Consumption | 35VA |
Size | 145*120*100mm |
Net Weight | 1.4KG |
Power Supply | AC:100-240V 50/60Hz;DC:DC12V±1.2V |
Battery | Li-Polymer 7.4V 1900mAh,more than 5 hours at 25ml/h |
Operating Conditions | Temperature:5°C-40°C Relative humidity:10-95% Atmosphere pressure:86kPa~106kPa |
Storage Conditions | Temperature:20°C-45°C Relative humidity:10-85% Atmosphere pressure:50.0kPa~106.0kPa |
Transport Conditions | Temperature:20°C-45°C Relative humidity:10-85% Atmosphere pressure:50.0kPa~106.0kPa |
Brand: HAWKMED
Additional Info | ||
---|---|---|